Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the wordpress-seo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /var/www/html/wp-includes/functions.php on line 6260 Vyaire's bellavista Ventilator Recall Deemed Class 1 - Medical Product Outsourcing
OEM News

Vyaire’s bellavista Ventilator Recall Deemed Class 1

Software configuration issues may cause ventilators to malfunction or stop.

The U.S. Food and Drug Administration (FDA) has classified Vyaire Medical’s December 23 recall of bellavista 1000 and 1000e ventilators as a Class I recall. 2,605 devices distributed from January 26, 2017 to December 23, 2021 were subject to the recall.
 
Thus far there have been 18 complaints and seven injuries reported. No deaths have been reported.
 
Vyaire recalled the ventilators that had specific software versions and configurations after reports the ventilators failed to ventilate and sounded a technical failure alarm. Devices with software version 6.0.1600.0 or higher may have a memory conflict between tasks when the data communication port is set to “HL7,” initiating technical failure alarm 305. The ventilator may then malfunction or stop.
 
Customers were urged to immediately disable the HL7 configuration data communication port until updated software is available to solve the issue.
 
The company released a voluntary field correction for the ventilators earlier this month.